56
Maintenance and care
AZURA® Detector DAD 6.1L, DAD 2.1L, MWD 2.1L Instructions, V6700
Process
Step
Figure
4.
Using the screwdriver, tighte
the two screws
4
or
7
on
the lamp socket.
5.
Connect the plug
5
or
6
and tighten the lock ring.
7
6
4
5
Fig. 38
Installing the deuterium
lamp
6.
Attach the lamp cover and fix
it with the screws
8
and
9
.
8
9
Fig. 39
Lamp cover
Bring the device into operation.
Note:
You can clean the lamp thoroughly with a lint free cloth and
isopro panol.
9.6 Checking the fittings
Check if all fittings are tight. In case you find fittings that are not tight,
tigh ten them up.
Capillary Screw Fittings
Torque
Stainless-steel fittings
5 Nm
PEEK fittings
0.5 Nm
9.7 Removing a leak
If the leak tray is filled with liquid, the leak has to be removed.
Cloth for drying the leak sensor
Process
1.
Remove the leak.
2.
Dry the leak tray with the cloth.
3.
Confirm the error message via Mobile Control, the software or the
standby switch (without software control).
Bring the device into operation.
Next step
Prerequisites
Auxiliary material
Next step