Maintenance and care
AZURA® Detector DAD 6.1L, DAD 2.1L, MWD 2.1L Instructions, V6700
Process
Step
Figure
4.
Using the screwdriver, loosen
the two screws
5
or
6
.
5.
Remove the lamp.
5
6
Fig. 35
Screws on the lamp socket
Insert a new lamp.
9.5.2 Installing the deuterium or halogen lamp
Removing the lamp cover switches off the device.
The device is switched off.
The lamp cover is removed.
The old lamp is removed.
Allen screwdriver, 2.5 mm
Performance decrease
Damage to the lamp and inaccurate measuring results due to residue
pos sible.
Do not touch the glass body with your hands.
Wear gloves.
Use a clean, soft cloth for cleaning.
Process
Step
Figure
1.
Hold the lamp at the lamp
socket and guide the glass
body into the lamp pod
1
.
2.
The bolt
2
of the deuterium
lamp sits in a notch in the
lamp socket.
1
2
Fig. 36
Lamp pod deuterium lamp
3.
Insert the halogen lamp
3
into the lamp pod at a slight
angle.
Fig. 37
Lamp pod halogen lamp
Next step
Prerequisites
Tools