32
Replacing the leak sensor
AZURA® CT 2.1 Service Instructions VSM-020
8. Replacing the leak sensor
8.1 Removing the leak sensor
The device has been switched off.
The power plug has been removed.
Torx screwdriver T10
Round rod with the size of the leak sensor
Hammer
Process
Figure
1.
Open the door of the device.
2.
Gently push the leak sensor
through the hole in the back
of the device by using the
round rod and the hammer.
Fig. 47:
Inside view of the device
Prerequisite
Tools
Electronic defect
Performing maintenance tasks on a switched on device can cause da-
mage to the device.
Î
Switch off the device
Î
Pull the power plug.
Electronic defect
Electrostatic discharge can destroy the electronics.
Î
Wear a protective bracelet against electrostatic discharge and ground.