26 Operation
AZURA® Conductivity Monitor CM 2.1S Instructions, V6710
Process
Display
3.
Read the measured value and modify the
% value of the factor until the measured value
and the target value match.
Press both arrow keys.
Hold down the left arrow key and use the
right arrow key to scroll until the correct
display appears.
Let go of the arrow keys and press the left or
right arrow key to select a value.
Example
MEASURED VALUE
SETTINGS
Target [mS/cm]
Actual [mS/
cm]
Deviation [%]
Factor [#] Input [mS/cm] Input [%]
50
50
0
1
50
100
100
94
-6
2
100
106
150
152
+1
3
150
99
299
299
0
4
299
100
Fig. 15:
Example of a fitted calibration function (black) in deviation
from the factory setting (green)
All factors = 100 %
Note:
If the adjustment of the calibration points for a factor # requires
more than 12 % deviation, the CM flow cell is contaminated and must be
flushed (see chapter 8.3) or replaced.
Enter additional settings or start the analysis.
5.7 Specifying the offset value
If the measured values consistently have an offset in the same direction,
these can be corrected with the
Offset value.
50
50
100
150
Actual (monitor display)
100
150
Target
Next steps