36
Transportation and storage
AZURA® Conductivity Monitor CM 2.1S Instructions, V6710
9.2 Disassembling the device
9.2.1 Disconnecting the power supply
The device has been switched off.
Process
1.
Pull the power supply out of the socket and afterwards remove the
plug from the device.
2.
Package the power supply and store it nearby the device.
9.2.2 Removing the CM flow cell
The CM flow cell must be disassembled before transport or storage.
To disassemble the CM flow cell, proceed as described in chapter 8.4.
Pack the device and the flow cell for transport or storage and store the
flow cell under the conditions described in chapter 9.3.1.
9.2.3 pH-Monitor abbauen
Das Gerät wurde ausgeschaltet.
Der Netzstecker wurde gezogen.
Process
Figure
1.
Disconnect the BNC plug con-
nection of the pH monitor to
the conductivity monitor and
provide connection
1
with
BNC blind plug.
2.
Disconnect the capillaries and replace the pH flow cell with the cou-
pling in the accessories kit of the pH monitor in the flow path
of the system.
3.
Detach the pH electrode from the pH flow cell and immerse it in a
KCl solution. The solution can be in a beaker or in the protective cap
of the pH electrode.
4.
Rinse the pH flow cell with water and store it dry.
9.3 Storing the device
9.3.1 Storage of the conductivity monitor and the CM flow cell
The functionality of the device remains warranted, if you select a storage
loca tion according to the following ambient conditions:
Prerequisites
Next steps
Prerequisites
1