DAD 6.1L/DAD 2.1L/MWD 2.1L user manual V6700, version 3.2
Optimizing the Detector
27
Default Settings
Using the Mobile Control or the Control Unit, you can re-set the detector to
its default settings.
Optimizing the Detector
In this chapter, you find information on how to optimize your detector.
Location
In order to ensure thermo stability and to prevent drift effects, note the fol-
lowing aspects:
Protect the detector against strong ventilation.
Protect the detector against direct sunlight.
Note the space requirements.
Avoid vibrations.
green
The deuterium lamp is
active.
Right LED
green
The device has been
switched on.
flashes green
The device not ready
for operation.
Wait until the
device is ready for
operation.
blue
The device is in
standby
Press the standby
button to end the
standby.
Color
Status
Operation
Parameter
Setting
Network
LAN DHCP, port 10001
Lamps
Deuterium ON, Halogen OFF (DAD 6.1L
only)
Time constant
2 s
Channels
Channel 1: WL = 254 nm, BW = 8 nm
Reference correction
Reference channel enabled for channel 2,
WL = 360 nm, BW = 30 nm
Extended linear
range
OFF
Analog out
Offset 1 = 0.00 mV, Scale 1 = 1 AU/V
Flow cell
Test cell
Event check
All events deactivated (o)
Date/Time
Current date/time
Wake-up
Current date/time
Leak sensor
ON, Sensitivity = low