DAD 6.1L/DAD 2.1L/MWD 2.1L user manual V6700, version 3.2
Installation
23
Process
1. Place the terminal strip
3
on a suitable sur-
face.
2. Push the depressor tool
1
into the opening
on the upper side.
3. Continue pushing the depressor tool down
and lead the cable
2
into the front end of
the terminal strip.
4. Remove the depressor tool.
5. Check whether the cables are tightly
attached.
6. Plug the terminal strip onto the multi-pin
connector.
Next steps
Finish the installation and perform the initial startup.
Integrator Connector
The integrator output supplies the current signal value as an analog voltage
(max. 5 V). The integrator output is connected to other instruments with
the analog connection cable or other special connection cables.
For details on the integrator connector, see the list below:
non-bipolar
4 channel
0 to 5 V
DAC 16 bit
scalable
Adjustable to offset
Power Connection
Note:
The nominal capacity of the connected devices must be maximum
50 % of the power supply to account for larger inrush currents when
switching on the modules.
The maximum power input is 75 W.
Prerequisite
The electrical power supply at the installation site must be connected
directly to the nearest main power line.
The power must be free from ripple, residual current, voltage peaks and
electromagnetic interference.
The connectors for the mains voltage are grounded accordingly.
The device receives sufficient power with reserve capacity
The device is intended for use with AC power networks of 100–240 V.
Power cable
Only the supplied power cable is to be used to connect the device to the
mains supply. Replace defective power cables only with accessories from
KNAUER. Only use power cables with a permission for use in your country.
In case of queries contact the Technical Support.
Electronic defect
Damage to the electronics if the device is turned on while connecting or
interrupting the power connection.
Switch off the device beforehand.
Power plug
Make sure that the power plug on rear of the device is always accessible, so
that the device can be disconnected from the power supply.
Next steps
Finish the installation and connect the device to the mains.
1
2
3