Ultrasonic Level Meter
Operation Manual
Extended connecting pipe for solid medium measurement
4.3.3Generation of false echo shall be avoided during the installation.
4.3.3.1 Devices and installation in the container
During sensor installation, it shall be ensured that the ultrasonic wave
beam is not blocked by other devices or feedstock. The bulges on the plane or
stair-like barriers in the container will have a great impact on the measurement
and a deflector can be provided on the embossment to reflect the false echo so
as to ensure accurate measurement.
Stair-like barriers in the container – inclined deflector needed to reflect the
false echo