▲Installation on the tank, pool, cover plate and support is basically the
same as above.
★After probe installation, the probe emission sur face must be exposed
from the cover plate or waveguide and it shall not be in the cover plate or
waveguide.
4.2.5 Liquid measurement
4.2.5.1 Flat-top tank
Normally, the flat-top tank has a short connecting pipe whose datum plane
is the undersurface of flange. Under the premise that the connecting pipe length
is ≤60mm, inner diameter is ≥100mm and inner wall is smooth and free of burr
and bulges, the measurement can be carried out if the emitting surface of
installed probe is 3cm below the flange undersurface.
Flange installation in a short connecting pipe
08.Transducer installed