Side 19
Ultrasonic Level Meter
Operation Manual
Probe installation on nipple joint
Similarly, the separate type sensor can be installed via top hoisting thread
and the dimensions of hoisting thread include M30
×
1.5, M32
×
1.5 and M38
×
1.5.
Hoisting thread connection at the separate type probe top
4.2.5.2 Arch tank top
For arch tank, it’s better not to install the meter in the middle of the tank
top. Instead, the meter shall be installed at 1/2 or 2/3 of the tank top radius
(under the premise that certain distance from the tank wall is met). The arch
tank top is like convex lens to the ultrasonic pulse. If the probe is installed at