Primary water pump
Fully installed and plumbed-in water pump Type Grundfos CR10-5 (see technical specifications and appendix).
The pump is assembled with an 3,0kW Motor from the manufacturer VEM.
(see technical specifications item 11.12.1).
The pump is switched on by turning the master switch to position "Auto"/”Hand” and the control switch for the
pump likewise to position "Auto"/”Hand”!.
The water is kept in constant circulation. The water pump is interlocked with the compressor, i.e. cooling is
unable to take place if the pump is not on.
It is essential that the pump is filled with the cooling medium and de-aerated before it is started.
Operating instructions CR
Multistage centrifugal pumps
1.General