Operating Instruction
Healthcare chiller of the KSC Series
Version „Revision 09“
Type KSC 215-L-U/S
page 24von138
\\FILER02\Allgemein\02 - Technische Unterlagen KKT\Originale\830xxx Bedienungsanleitungen\KSC\83002302.Ka KKT chillers_KSC215-US.doc
7
Notes on the water connections
The water connections between industrial coolers of the KSC series and the IFP of Siemens can be made of
stainless steel, copper or plastic. The nominal widths of the piping for distances of up to approx. 25 m should
match at least the dimensions of the fittings on the appliance. (see technical specifications)
Reductions in diameter should be avoided. In case of longer pipelines, the pump pressure should be tested.
When choosing the pipe materials, ensure that no electrochemical series are created.
The proper way to seal the European fittings can be done with any of following steps.
1.
Pipe sealing cord. There are a number of brands available, however we use Loctite 55
2.
Teflon Tape and a Anaerobic sealant.
3.
Teflon Tape an Nylog sealant.
As with any sealant, the application instruction must be followed for proper use.
See also enclosed “ Installing Instruction “