66
Combilog 1020+
U 3.xx
Date: dd.mm.yy
Time: hh.mm.ss
LOG: 0 253kb
free: 27d
PCMCIA: 1904kb
free: 100d
Chan 1
xx [units]
Chan x
xx [units]
Service mode
normal
Weather station
Adr. : 1 19200N
Auto. off: OFF
LEDs: ON
Scan Rate: 4s
Averaging: 10m
Display Contrast
adjust?
Input
disable
Combilog 1020+
U 3.xx
Figure 42: Main menu (left) and Configuration menu (right) of COMBILOG data logger.
Step 1: Enable the press/rotary knob of the data logger
For safety reasons the press/rotary knob of the dat alogger is disabled (automatically done by the
COMBILOG data logger), meaning that part of the configuration menu cannot be accessed. In order to
enable this knob, go to “
COMBILOG 1020+
” of the main menu by rotating the press/rotary knob. Press
the knob briefly. The configuration menu will be accessed. Rotate the knob until “
Input disabled
”
appears on the display (see Figure 42). Again press the knob briefly. Rotate the knob again until
“
Input < enable >
” appears and press the knob briefly. The knob is now enabled to modify all settings
of the data logger manually. Go back to the main menu, by pressing the knob for about 1 second, you
will return to the main menu (“
COMBILOG 1020+
” appears).
Note that the knob will be disabled again after 30 seconds if no operations are performed on the knob.
Input
< disable >
Input
< enable >
Input
disable
Figure 43: Lock/Unlock of the press/rotary knob of the COMBILOG data logger.
Summary of Contents for LAS
Page 1: ...Instruction Manual Large Aperture Scintillometer LAS...
Page 28: ...27 Figure 17 Operational weather station of LAS RET system...
Page 29: ...28...
Page 39: ...38...
Page 45: ...44...
Page 49: ...48...
Page 51: ...50 Figure 33 Configuration screen and default settings of supplied RF modems...
Page 55: ...54...
Page 69: ...68...
Page 71: ...70 APPENDIX 9 CONNECTION PLAN LAS BET LAS RET Figure 47 Connection plan LAS BET system...
Page 72: ...71...
Page 73: ...72 Figure 48 Connection plan weather station LAS RET system...
Page 74: ...73 Figure 49 Connection plan scintillometer LAS RET system...