The analyser backs to the list of present operators.
➢
Press
Esc
to quit the screen.
4.6.2. Modify the operator name
“Settings”
screen is displayed
.
➢
Go to
“Operator”
with the rotative button then press
OK
.
➢
Press the function key
.
The keyboard appears.
➢
Modify the name with the rotative button : press
OK
to validate.
➢
Press the function key
to validate the operator name.
The analyser backs to the list of current operators.
➢
Press
Esc
to quit the screen.
4.6.3. Delete an operator
“Settings”
screen is displayed
.
➢
Go to
“Operator”
with the rotative button then press
OK
.
➢
Go to the operator to delete with the rotative button.
➢
Press the function key
.
➢
Press
Esc
to quit the screen.
4.7. Reset the Factory settings of the analyser
It is possible to reset set the analyser with its factory settings.
“Settings”
screen is displayed
.
➢
Press the function key
.
A message is displayed and asks to confirm the return to factory settings.
➢
Go to
“OK”
with the rotative button then press
OK
.
To cancel this return to factory settings when the
“Back to factory setting”
message is displayed :
➢
Press
Esc
or go to
“Cancel”
with the rotative button then press
OK
.
4.7.1. Units and values of measurement settings of the factory settings
•
Temperature :
°C
•
Draft :
Pa
•
Differential pressure :
mbar
•
CO :
ppm
•
CO
2
:
%
•
NO :
ppm
•
NO
2
:
ppm
•
NOx :
ppm
•
SO
2
:
ppm
•
CH
4
:
ppm
•
CO (O
2
) :
mg/m
3
•
NO (O
2
) :
mg/m
3
•
NOx (O
2
) :
mg/m
3
Alarm thresholds :
•
O
2
:
3.0%
•
CO :
100 ppm
•
CO amb :
20 ppm
•
CH
4
:
1200 ppm
•
Excess air (λ) :
1.10
•
Flue gas temperature (Tf) :
250.0°C
Reference :
•
O
2
:
21.0%
•
Altitude :
800 m
•
Atmospheric pressure :
1013.25 hPa
CO protection threshold :
2000 ppm
Duration of autozero and purge
: 1 minute
Created operators and combustibles will be deleted. Created customers and boilers and inspections on boilers are not deleted after a return to
factory settings.
Set the different parameters of the analyser