19
2.2
Display and Modification of Alarm Set Value (AL)
** In the normal display screen, press
for 3 seconds to enter the display and modification of the alarm set value
Display Screen
Default
Screen Name
Modify Parameters and Process Description
Alarm 1 set
value (AL1)
1. Setting example:
If you want to display 50.0, AL1 alarm action.
Here AL1 must be set to 50.0.
2. Press
to enter parameter modification mode, the value will
flash.
3. Press
or
to modify the set value of the alarm.
Modifiable range -199999 ~ 999999
4. After setting is completed, press ENT to save the modified
parameters and enter the next parameter setting page.
Alarm 2 set
value (AL2)
2.3
Basic Unit Setting Group
** In the normal display screen, press
for 3 seconds to enter the display value setting group
Display Screen
Default
Screen Name
Modify Parameters and Process Description
Display zero
point deviation
correction value
(doFSt)
1. Zero point correction example: input 0V, the rated display is
0, the actual display is 3, input 3 here to correct the deviation.
Modifiable range: -199999 ~ 999999.
2. Press
to enter parameter modification, press
or
to
modify the value.
3. Press ENT to save the modified parameters and enter the
next parameter setting page.
Display
deviation
correction value
(dGAin)
1. Display correction example: input 10V, the rated display is
100.0, the actual display is 99.8. Actual display = dGAin,
100.0, 99.8 = 1.002, here you need to set 1.002.
2. Press
to enter parameter modification, press
or
to
modify the value.
3. Press ENT to save the modified parameters and enter the
next parameter setting page.
Basic unit
decimal point
position setting
(dP)
1. Set basic unit decimal point position: display 100.0 needs to
be changed to 10.00. Here change the original 1 to 2.
Modifiable digits: 0, 1, 2, 3, 4, 5 (digits)
2. Press
to enter parameter modification, press
or
to
modify the value.
3. Press ENT to save the modified parameters and enter the
next parameter setting page.
Lowest display
setting value
(dSPL)
1. Setting example: input 0V and display the lowest 10, here
you need to set it to 10. Modifiable range: -199999 ~ 999999.
2. Press
to enter parameter modification, press
or
to
modify the value.
3. Press ENT to save the modified parameters and return to the
display value setting page.
* Press the Z key on this page to correct the minimum input
signal. (Suitable for load cell)
Forward 2nd
point display
setting value
(r.dSPM)
1. Setting example: LOAD CELL inputs signal at the forward
2nd point shall display 500. Set it to 500 here. Modifiable
range: -199999 ~ 999999.
2. Press
to enter parameter modification, press
or
to
modify the value.
3. Press ENT to save the modified parameters and return to the
display value setting page.
* Press the Z key on this page to correct the forward 2nd point
input signal. (Suitable for load cell)