26
Alarm start
delay time
setting value
(Sdt)
Setting description:
1. When the displayed value reaches the alarm startup
delay setting range, alarm will not operate until the set time
has passed. (P.S.: This function must be used with Sb.)
Modifiable range: 0 ~ 99 (second)
2. Press
to enter parameter modification, press
or
to modify the value.
3. After setting is completed, press ENT to save the
modified parameters and enter the next parameter setting
page.
Warning!
KTM-IWT50 maximum load torque upper limit is 50N.m, KTM-IWT400 maximum load torque upper limit is
400N.m. If exceeding the upper limit, an alarm sound will be issued. If it exceeds the load for a long time, it will
cause damage to the tester and the value will be unstable and difficult to adjust and we don’t warranty it.
3.3
Analog Output (AoP) Setting Group Flow and Display
** After input the correct password, press
again to select the alarm output setting group screen
Display Screen
Default
Screen Name
Modify Parameters and Process Description
Analog output
polarity setting
(PoLAr)
Setting description:
1. Select NO as positive output: 0 ~ 10Vdc
Select YES as the positive and negative output:
-10 ~ +10Vdc
2. Press
to enter parameter modification, press
or
to modify the value.
3. Press ENT to save the modified parameters and enter the
next parameter setting page.