TOS5300_INTERFACE
71
App
D
Command processing time
A certain amount of time is required before the commands shown in the following table are
received by the TOS5300 Series.
The processing times shown here are standard values, not guaranteed values.
The processing times vary depending on the settings and the measurement conditions.
The values shown below do not include hardware response times.
Command
USB processing time
(ms)
Description
SOUR:FUNC:MODE ACW
2
Sets the test mode
SOUR:VOLT 1kV
2
Sets the test voltage
SOUR:VOLT:TIM 1S
2
Sets the test time
SENS:JUDG 5mA
2
Sets the upper limit
MEAS:VOLT?
20
Starts a new measurement and queries
the voltage
STAT:OPER:TEST:COND?
3
Queries register status