AC Withstanding Voltage Test (ACW) Conditions
3
Co
m
m
a
n
d
R
e
fe
re
n
ce
Test time setting (ACW)
SOUR:VOLT:TIM
Sets the test time (TIMER). This setting is enabled when SOUR:VOLT:TIM:STAT is set to ON.
Command
SOURce[:ACW]:VOLTage:TIMer {<numeric>|MIN|MAX}
SOURce[:ACW]:VOLTage:TIMer? {MIN|MAX}
Parameter
Value:
0.1 to 999.0
(The default value is 0.1.)
Unit:
S
Response
Returns the test time in <NR3> format.
SOUR:VOLT:TIM:STAT
Sets whether to stop testing after the set test time elapses (TIMER ON/OFF). Use SOUR:VOLT:TIM to
set the test time.
Command
SOURce[:ACW]:VOLTage:TIMer:STATe {ON|OFF|1|0}
SOURce[:ACW]:VOLTage:TIMer:STATe?
Parameter
Value:
ON (1)
Testing is stopped after the test time elapses (default).
OFF (0)
Testing is not stopped after the test time elapses.
Response
Returns whether testing is stopped after the test time elapses in <NR1> format.
Start voltage setting (ACW)
SOUR:VOLT:STAR:STAT
Sets whether the start voltage is used. The start voltage is 50 % of the test voltage.
Command
SOURce[:ACW]:VOLTage:STARt:STATe {ON|OFF|1|0}
SOURce[:ACW]:VOLTage:STARt:STATe?
Parameter
Value:
ON (1)
The start voltage is used.
OFF (0)
The start voltage is not used.
Response
Returns whether the start voltage is used in <NR1> format.