44
TOS5300_INTERFACE
Various Settings
Calibration period setting
SYST:CONF:CAL:DUE:CONT
Set the number of months until the calibration period.
Command
SYSTem:CONFigure:CALibration:DUE:CONTrol
{<numeric>|MIN|MAX|INFinity}
SYSTem:CONFigure:CALibration:DUE:CONTrol? {MIN|MAX}
Parameter
Value:
3 to 36 (The default value is 12)
INFinity
Indefinitely.
Response
Returns the number of months until the calibration period in <NR1> format.
Unit:
month
Other settings
SYST:ERR
Queries the oldest error or event from the error/event queue. The error/event queue can hold up to
255 errors.
The error/event queue is cleared if a *CLS command is sent.
Command
SYSTem:ERRor[:NEXT]?
Response
Returns the oldest error or event from the error/event queue in the following format.
Example
If the queue is empty,
this command returns 0 "No error."
Example
If a command has been received that cannot be executed in the current operating state
of the TOS5300 Series,
this command returns -221, "Settings conflict."
If there are multiple errors in the queue, the errors are returned from the oldest error first. Errors are
cleared from the queue after they have been returned.
SYST:KLOC
Locks or releases panel operations (key lock). When keys are locked, only the START and STOP
switches are enabled.
Command
SYSTem:KLOCk {ON|OFF|1|0}
SYSTem:KLOCk?
Parameter
Value:
ON (1)
Panel operations are locked.
OFF (0)
Locked panel operations are released (default).
Response
Returns the key lock status in <NR1> format.