TOS5300_INTERFACE
41
Querying Measured Values (SEQuence1:ACQuire)
3
Co
m
m
a
n
d
R
e
fe
re
n
ce
RES
Queries the results of the previous test.
Command
RESult[:IMMediate]?
Response
Returns the following results of the previous test: test number, program number, test mode, test
start time, voltage, current, resistance, test time, and judgment result. Each result is separated by a
comma in the following format: <NR1>, <NR1>,<character>, <NR1>, <NR3>, <NR3>, <NR3>,
<NR3>, <character>.
Response format <NR1>Test number
A counter that is incremented each time a test is executed. After 4294967295, the
count returns to 0.
<NR1>Program number
For a single test, 1 is always returned. For an auto test, 1 is returned for the first test,
and 2 is returned for the next test.
(Example) ACW and then IR
If you send a RES? query after the ACW test, 1 is returned. If you send the query after
the IR test, 2 is returned.
<Character>Test mode {ACW|DCW|IR}
ACW: AC withstanding voltage test
DCW: DC withstanding voltage test
IR: Insulation resistance test
<NR1>Test start time yyyy, mm, dd, hh, mm, ss (year, month, day, hour, minute, second)
<NR3>Voltage
Measured voltage
<NR3>Current
Measured current
If the test result is fail, the judgment criterion is returned.
<NR3>Resistance
Measured resistance
If the test result is fail, the judgment criterion is returned.
<NR3>Test time
The test time is returned.
<Character>Judgment result {PASS|U-FAIL|L-FAIL|PROT|ABORT}
PASS: The test was successful.
U-FAIL: A value exceeding the upper limit was detected.
L-FAIL: A value less than the lower limit was detected.
PROT: A protection function was activated, and the test was stopped.
ABORT: Test was aborted.