PFX2000
Appendix A-9
■
Moving average process
You can decrease the
fl
uctuation of the measured value even further by performing
moving average. The unit allows you to select the average count from n = 0 (OFF),
2, 4, and 8. Fig. A-13 shows an example of the moving average process when n is
set to 4.
Fig. A-13
When the Average Count is Set to 4
By default, the average count is set 0 (OFF). For details on setting the moving aver-
age, see section 4.3.1, “Module” in the
BPChecker2000 User’s Manual
.
• The moving average also functions during pulse charge/discharge operation. Note
that the updating of the measurement result takes longer, if the moving average is
enabled when the pulse time width is set large.
Measured value
Moving average result
Average period 1
Average period 2
Average period 3
Number of measurements
NOTE