5-6 Specifications
PFX2000
Other functions
Item
PFX2011
PFX2021
Range switching
Current range
High
0.0 mA to 5000.0 mA
–
Low
0.00 mA to 500.00 mA
Power saving function
Discharge power regeneration
–
Regeneration operation in the
battery voltage range speci
fi
ed
for discharging.
Main power short circuit mode
–
Short-circuits the main power
supply when the battery voltage
is high during discharge.
Vce control
Controls to keep the loss at the electronic load section constant.
Fan control
Turns ON at approximately 45
°
C or more (fast), OFF at 40
°
C or
less (slow).
Standby mode
Shuts off the main power supply
input when in standby mode.
Stops the main power supply
operation.
Auto
fi
ne function
Correctable
range
±
30 digits
(Enable/disable selectable. Normally disabled. During constant cur-
rent.)
Update rate
120 ch/1 s Max
(when using two USB ports (two
control units) with 30 channels
per port)
60 ch/1 s Max
(when using two USB ports (two
control units) with 15 channels
per port)
Expansion function
DC impedance measurement of batteries.
(when the PFX2211 is connected)
Status monitoring
Standard State
Power OFF, Idle, Idle (DUT Connection Warning), Idle (Protection
Warning) , H Protection, Charge, Charge Rest, Charge SYNC,
Charge End, Discharge, Discharge Rest, Discharge SYNC, and Dis-
charge End
Operate
OFF, CC, CV, CP, Power Limit, and CC Pulse