4-8 Maintenance
PFX2000
■
Symptom 8: Correct temperature is not indicated.
■
Symptom 9: After the test is started, the ALARM/WARNING
LED blinks in red.
Symptoms related to the control unit
■
Symptom 10: The POWER LED does not illuminate even if the
control unit is connected to the PC via the USB.
■
Symptom 11: The MON ERR LED remains illuminated.
Check Item
Cause and Remedy
The readout error is large.
• The readout error will be large, if you are not using the
speci
fi
ed thermistor or if other components such as a
capacitor is attached to the thermistor.
Use only the speci
fi
ed thermistor.
• Check the contact between the DUT and the thermistor.
Even a slight gap will greatly affect the measurement
result.
Check Item
Cause and Remedy
An alarm indication is active.
• The charge/discharge power supply unit has detected
some kind of error and the test was aborted and is in an
alarm status.
Correct the problem by referring to section A.2, “Cause
of Alarms and Their Correction” in the
BPChecker2000
User’s Manual.
Check Item
Cause and Remedy
The PC is not turned on. Or, the OS
is booting up.
• The control unit receives power from the USB. Thus,
turn on the PC. Also, wait until the OS boots up com-
pletely.
• The USB port may not be properly identi
fi
ed on the PC.
See section 5.2.1, “If the Test Executive Cannot Iden-
tify Controllers” in the
BPChecker2000 User’s Man-
ual
.
The self-powerd USB hub is not
turned on.
• Turn on the hub.
Check Item
Cause and Remedy
An error occurred in the communi-
cation control of the application
software.
• Check whether the application software is running nor-
mally.
If there is a problem, restart the application software.
You may need to restart your PC. When the communi-
cation control by the application software returns to
normal, the MON ERR LED turns off.