PFX2000
Appendix A-5
PFX2021
As shown in Fig. A-7, voltage measurement and current measurement are per-
formed independently at the same time regardless of the speci
fi
ed pulse time width.
Fig. A-7
Example 3 during Pulse Charge
■
Pulse voltage measurement
PFX2011 and PFX2021
The following two types of measurement modes are available for pulse voltage
measurement.
1. Peak measurement
Measures the maximum and minimum voltages in one pulse cycle. If a pulse cur-
rent with a short time width is repeated, this measurement method is often used.
2. Point measurement
Measurement is performed according to the speci
fi
ed measurement points (2
points). This is useful when you wish to arbitrary
fi
x the measurement point.
• If the battery voltage
fl
uctuation is large with respect to the change in the pulse
current as shown in Fig. A-8, the measured voltage returns an average value
within the conversion period.
Fig. A-8
When the Voltage is Fluctuating during the Measurement Period
Pulse current
waveform
Battery voltage
waveform
Voltage measurement point
Pulse current integration
Current large
↑
Current small
↓
I1
I2
I3
I1
I2
I3
≈
230
µ
s
NOTE
Pulse current waveform
Battery voltage waveform
The measured value is
the average value within
the measurement period.