2
PAT-T
SOURce subsystem
SCPI Command
Setting
Default
Resp. *RST
Description
R/W
†
Program header
Parameter
Unit
[SOUR]
:VOLT
[:LEV][:IMM][:AMPL]
numeric
0 % to 105 %
of the rated output
voltage
V
0
NR3
Yes
Sets the voltage.
R/W 1
:EXT:SOUR
char
NONE | VOLT |
RES | IRES |
FVOL
*1
NONE
char
Yes
Sets the CV control
mode setting.
R/W 3
:FINE
*2
NRf
-10000 to 10000
0
Yes
Sets the fine
adjustment setting
for voltage.
W
3
:LIM
[:AMPL]
V
NR3
Yes
Queries the voltage
limit setting.
R
3
:AUTO
bool
OFF
NR1
Yes
Enables/Disables the
voltage limit setting.
R/W 3
:PROT
[:LEV]
numeric
10 % to 111.5 %
of the rated output
voltage
V
111.5 % of
the rated
output
voltage
NR3
Yes
Sets the OVP.
R/W 1
:TRIG
numeric
0 % to 105 %
of the rated output
voltage
V
0
NR3
Yes
Target voltage using
a trigger.
R/W 1
:CURR
[:LEV][:IMM][:AMPL]
numeric
0 % to 105 %
of the rated output
current
A
105 % of the
rated output
current
NR3
Yes
Sets the current.
R/W 1
:EXT:SOUR
char
NONE | VOLT |
RES | IRES |
FVOL
NONE
char
Yes
Sets the constant
current control mode. R/W 3
:FINE
NRf
-10000 to 10000
0
Yes
Sets the fine
adjustment setting
for current.
W
3
:LIM
[:AMPL]
A
NR3
Queries the current
setting limit.
R
3
:AUTO
bool
OFF
NR1
Yes
Enables/Disables the
current setting limit.
R/W 3
:PROT
[:LEV]
numeric
10 % to 111.5 %
of the rated output
current
A
111.5 % of
the rated
output
current
NR3
Yes
Sets the OCP.
R/W 1
:TRIG
numeric
0 % to 105 %
of the rated output
current
A
105 % of the
rated output
current
NR3
Yes
Target current using
a trigger.
R/W 1
*1.
FVOL parameter is available for 8 kW type products only.
*2.
The command is available for 8 kW type products only.
PAT-T series Communication Interface Manual
KIKUSUI ELECTRONICS CORP.