Calibration Procedures
1
Keysight U1281A/U1282A Service Guide
49
Capacitance
Open
oPEn Open
terminals
10.00 nF
3.00 nF
0.9 to 1.1 x Reference value
10.00 nF
0.9 to 1.1 x Reference value
100.00 nF
10.00 nF
0.9 to 1.1 x Reference value
100.00 nF
0.9 to 1.1 x Reference value
1.00
μ
F
0.10
μ
F
0.9 to 1.1 x Reference value
1.00
μ
F
0.9 to 1.1 x Reference value
10.00
μ
F
1.00
μ
F
0.9 to 1.1 x Reference value
10.00
μ
F
0.9 to 1.1 x Reference value
100.00
μ
F
10.00
μ
F
0.9 to 1.1 x Reference value
100.00
μ
F
0.9 to 1.1 x Reference value
1.00 mF
0.10 mF
0.9 to 1.1 x Reference value
1.00 mF
0.9 to 1.1 x Reference value
10.00 mF
1.00 mF
0.9 to 1.1 x Reference value
10.00 mF
0.9 to 1.1 x Reference value
100.00mF
1.00 mF
0.9 to 1.1 x Reference value
10.00 mF
0.9 to 1.1 x Reference value
Diode
Short
SHort
Zero reference point
3.10 V
2.00 V
0.9 to 1.1 x Reference value
Short
SHort
Short V
Ω
/ COM terminals
DC
μ
A/mA
Open
oPEn
Open terminals
60.00
μ
A
30.00
μ
A
0.9 to 1.1 x Reference value
600.00
μ
A
300.00
μ
A
0.9 to 1.1 x Reference value
6.00 mA
3.00 mA
0.9 to 1.1 x Reference value
60.00 mA
30.00 mA
0.9 to 1.1 x Reference value
600.00 mA
300.00 mA
0.9 to 1.1 x Reference value
Table 1-19
Adjustment input values (continued)
Test function
Step
Reference value
Valid reference input