Calibration Procedures
26
Keysight U1281A/U1282A Service Guide
Calibration Process
1
Prior to performing the verification tests, see the
2
Perform the verification tests to characterize the multimeter; see
3
Unsecure the multimeter for calibration; see
4
Prior to performing the adjustments, see the
5
Perform the adjustment procedure; see
6
Secure the multimeter against unauthorized calibration; see
on page 54. Ensure that the multimeter has quit the
adjustment mode and is turned off.
7
Record the new security code and calibration.
Test Considerations
For optimum performance, all procedures should comply with the following
recommendations:
– The performance verification test or adjustment should be performed under
laboratory conditions where the ambient temperature can be controlled.
– The multimeter should be put under the laboratory environment for at least 1
hour.
– Ensure that the calibration ambient temperature is stable and is between 18 °C
and 28 °C. Ideally the calibration should be performed at 23 °C ± 1 °C.
– Ensure that the ambient relative humidity is less than 80%.
– Allow a warm-up period of 5 minutes.
– Use a shielded twisted pair of PTFE-insulated cables to reduce settling and
noise errors. Keep the input cables as short as possible. Long test leads can
also act as antennas which may pick up AC signals.
– Connect the input cable shields to earth ground.