Calibration Procedures
1
Keysight U1281A/U1282A Service Guide
43
To reset the calibration security code to its factory default
If you have forgotten the correct calibration security code, you may follow the
steps below to reset the calibration security code to the factory default code
.
1
Note down the last four digits of the multimeter’s serial number (located at the
bottom of the rear panel) before you begin.
2
Press
and
together for more than 1 second during normal
operation mode to enter unsecured mode.
and
will be shown on
the secondary and primary display.
3
Press
for more than 1 second to enter the Default Security Code (DSC)
mode.
and
will be shown on the secondary and primary display.
4
Enter the last four digits of the multimeter’s serial number.
– Press or on
to move the cursor to the left or to the right.
– Press
or
to increment or decrement the digit.
5
Press
to confirm your entry.
– If the four digits entered are correct, the display will show
. The
calibration security code is now set to its factory default code,
.
– If the incorrect security code is entered, an error code will appear briefly,
after which the multimeter will prompt you for the security code again.
– To enter a new security code, see
“To change the calibration security code”
on page 41. Ensure that you record down the new security code.
NOTE
If you do not have a record (or have lost the record) of the security code, first try
the factory default code,
.