Calibration Procedures
48
Keysight U1281A/U1282A Service Guide
Resistance
Short
SHort
Zero reference point
60.00 M
Ω
oPEn
Open terminals
10.00 M
Ω
0.9 to 1.1 x Reference value
6.00 M
Ω
3.00 M
Ω
0.9 to 1.1 x Reference value
600.00 k
Ω
300.00 k
Ω
0.9 to 1.1 x Reference value
60.00 k
Ω
30.00 k
Ω
0.9 to 1.1 x Reference value
6.00 k
Ω
3.00 k
Ω
0.9 to 1.1 x Reference value
600.00
Ω
300.00
Ω
0.9 to 1.1 x Reference value
60.00
Ω
30.00
Ω
0.9 to 1.1 x Reference value
Short
SHort
Short V
Ω
/ COM terminals
Table 1-19
Adjustment input values (continued)
Test function
Step
Reference value
Valid reference input