Performance Tests and Calibration
6
Keysight U1251B/U1252B User’s and Service Guide
149
Step
Test Function
Cal Range
Input
Cal Item
U1251B
U1252B
11
Press
button to go to
500 µA
30 µA,1 kHz
030.00 µA
030.00 µA
µA mode
300 µA,1 kHz
300.00 uA
300.00 µA
5000 µA
300 µA,1 kHz
0300.0 µA
0300.0 µA
3000 µA, 1 kHz
3000.0
µA
3000.0 µA
12
Turn the Rotary Switch to the
position
Open
Input terminal
open (remove any
test leads and
Short Plugs from
the input terminal)
oPEn
oPEn
50 mA
30 mA
30.000 mA
30.000 mA
440 mA
300 mA
300.00 mA
300.00 mA
Move the test lead from uA.mA and COM terminal to A and COM terminal
Caution: Connect the calibrator to hanheld multimeter's A and COM terminals before applying 3A and 10A
5 A
3 A
3.0000 A
3.0000 A
10 A
10 A
10.000 A
10.000 A
Move the test lead from A and COM terminal to uA.mA and COM terminal
13
Press
button to go to
50 mA
3 mA, 1 kHz
03.000 mA
03.000 mA
mA mode
30 mA, 1 kHz
30.000 mA
30.000 mA
440 mA
30 mA, 1 kHz
030.00 mA
030.00 mA
300 mA, 1 kHz
300.00 mA
300.00 mA
Move the test lead from uA.mA and COM terminal to A and COM terminal
Caution: Connect the calibrator to hanheld multimeter's A and COM terminals before applying 3A and 10A
14
Press
button to go to
5 A
0.3 A, 1 kHz
0.3000 A
0.3000 A
A mode
3 A, 1 kHz
3.0000 A
3.0000 A
10 A
3 A, 1 kHz
3.0000 A
3.0000 A
10 A, 1 kHz
10.000 A
10.000 A