Maintenance 5
Keysight U1251B/U1252B User’s and Service Guide
113
3
To test Fuse 1, place the tip of the test probe on the right
half of input terminal
. Ensure that the probe tip
touches the metal inside the input terminal, as shown in the
figure above.
4
To test Fuse 2, place and touch the tip of the test probe
on the right half of input terminal
. Ensure that the
probe tip touches the metal inside the input terminal.
5
Observe the reading on the instrument's display. Refer to
Table 5- 3
below for the possible readings that could
appear).
6
Replace the fuse when
OL
is displayed.
Table 5-3
Measurement readings for fuse checking
Current input terminal
Fuse
Fuse rating
Fuse OK (approximately)
Replace fuse
Displayed readings
1
440 mA/1000 V
6.2 M
Ω
OL
2
11 A/1000 V
0.06
Ω
OL