Maintenance 5
Keysight U1251B/U1252B User’s and Service Guide
109
6
The smart charging mode will start if the meter passes
the self- test. The charging time is limited to not more
than 220 minutes. The secondary display will count down
the charging time. During charging, no buttons can be
pressed. The error message may appear during charging to
alert the user of any overcharging of the battery.
C-Err
1
If charging battery more than 12 V or
less than 5 V
2
In 3 minutes, if the battery voltage
does not go upwards then charge
error
Table 5-2
Error messages (continued)
N O T E
•
If the
OL
message appears while the battery is inside, do not charge the battery.
•
If the
C-Err
message appears, check if the battery is the specified type. Use the
battery as specified in this manual. Ensure that the battery is the correct
rechargeable type before charging it again. After replacing with the correct
rechargeable battery, press Shift to perform the self-test again. If the C-Err
message reappears, replace with a new battery.