Performance Tests and Calibration
6
Keysight U1251B/U1252B User’s and Service Guide
129
Perform the verification test steps in the following
Table 6- 2
:
Table 6-2
Verification Test
Step
Test Function
Range
5520A Output
Error from nominal 1 year
U1251B
U1252B
1
Turn the rotary switch to the
position
[
1
]
5 V
5 V, 1 kHz
± 32.5 mV
± 22.5 mV
4.5 V, 10 kHz
± 169.5 mV
± 71.5 mV
4.5 V, 20 kHz
N/A
± 169.5 mV
4.5 V, 30 kHz
± 169.5 mV
N/A
4.5 V,100 kHz
N/A
± 169.5 mV
50 V
50 V,1 kHz
± 325 mV
± 225 mV
45 V,10 kHz
± 1.695 V
± 715 mV
45 V, 20 kHz
N/A
± 1.695 V
45 V, 30 kHz
± 1.695 V
N/A
45 V, 100 kHz
N/A
± 1.695 V
500 V
500 V, 1 kHz
± 3.25 V
± 2.25 V
1000 V
1000 V, 1 kHz
± 10 V
± 8.0 V
2
Press
button to go to
frequency mode
9.9999 kHz
0.48 V, 1 kHz
± 500 mHz
± 500 mHz
3
Press
button to go to
Duty Cycle mode
0.01% – 99.99%
5.0 Vpp @
50%, Square
Wave, 50 Hz
± 0.315%
± 0.315%