Performance Tests and Calibration
6
Keysight U1251B/U1252B User’s and Service Guide
133
Step
Test Function
Range
5520A Output
Error from nominal 1 year
U1251B
U1252B
1000.0
μ
F
1000.0
μ
F
± 10.5
μ
F
± 10.5
μ
F
10.00 mF
10.00 mF
± 0.105 mF
± 0.105 mF
100.00 mF
10.00 mF
± 0.4 mF
± 0.4 mF
14
Press
button to go to
mode
[
8
][
13
]
–200 ºC until
1372 ºC
0 ºC
± 3 ºC
± 3 ºC
100 ºC
± 3.3 ºC
± 3.3 ºC
15
Turn the rotary swich to the
position
500
μ
A
500
μ
A
± 0.55
μ
A
[
9
]
± 0.3
μ
A
[
9
]
5000
μ
A
5000
μ
A
± 5.5
μ
A
[
9
]
± 3
μ
A
[
9
]
16
Press
button to go to
µA mode
[
1
]
500
μ
A
500
μ
A, 1 kHz
± 4.2
μ
A
± 3.7
μ
A
500
μ
A, 20 kHz
± 15.8
μ
A
± 3.95
μ
A
5000
μ
A
5000
μ
A, 1 kHz
± 42
μ
A
± 37.0
μ
A
5000
μ
A, 20 kHz
± 0.156 mA
± 39.5
μ
A
17
Turn the rotary switch to the
position
50 mA
50 mA
± 0.105 mA
[
9
]
± 80
μ
A
[
9
]
440 mA
400 mA
± 0.93 mA
[
9
]
± 0.65 mA
[
9
]
18
Press
button to go to
mA mode
[
1
]
50 mA
50 mA, 1 kHz
± 0.42 mA
± 0.37 mA
50 mA, 20 kHz
± 1.56 mA
± 0.395 mA
440 mA
400 mA, 45 Hz
± 3.4 mA
± 3.0 mA
400 mA, 1 kHz
± 3.4 mA
± 3.0 mA
Caution: Connect the calibrator to handheld multimeter's A and COM terminals before applying 5A and 10A
5 A
5 A
± 16 mA
± 16 mA
10 A
[
10
]
10 A
± 40 mA
± 35 mA