Performance Tests and Calibration
6
Keysight U1251B/U1252B User’s and Service Guide
142
Valid adjustment input values
Adjustment can be accomplished using the following input
values below.
Table 6-3
Valid adjustment input values
Function
Range
Valid Amplitude Input Values
5 V, 50 V, 500 V, 1000 V
0.9 to 1.1 x Full Scale
(for U1251B)
5 V, 50 V, 500 V, 1000 V
0.9 to 1.1 x Full Scale
(for U1252B)
5 V, 50 V ,500 V, 1000 V
0.9 to 1.1 x Full Scale
50 mV, 500 mV, 1000 mV
0.9 to 1.1 x Full Scale
500
µ
A, 5000
µ
A
0.9 to 1.1 x Full Scale
50 mA, 440 mA, 5 A, 10 A
0.9 to 1.1 x Full Scale
Ω
500
Ω
, 5k
Ω
, 50 k
Ω
,500 k
Ω
, 5M
Ω
, 50 M
Ω
0.9 to 1.1 x Full Scale
Diode
0.9 to 1.1 x Full Scale
/
10 nF, 100 nF, 1000 nF, 10
µ
F, 100
µ
F, 1000
µ
F,
10 mF / 0
°
C
Make sure to provide 0
°
C with
ambient compensation