Installation Note N5242-90125 5
Description of the Upgrade
Getting Prepared
Getting Prepared
To successfully install this upgrade kit, you will need the following:
— A PDF copy or a paper copy of the PNA Service Guide - refer to
“Downloading the Online PNA Service Guide”
below.
— An ESD-safe work area - refer to
“Protecting Your Workspace from
below.
“Tools Required for the Installation” on page 7
— Enough time - refer to
“About Installing the Upgrade” on page 9
.
— Test equipment for the post-upgrade adjustments. To view the equipment
list, click the Chapter 3 bookmark "Tests and Adjustments" in the PDF
Service Guide
1
.
The PNA contains extremely sensitive components that can be ruined if
mishandled. Follow instructions carefully when making cable connections,
especially wire harness connections.
The person performing the work accepts responsibility for the full cost of
the repair or replacement of damaged components.
IMPORTANT!
Before you begin this upgrade:
—
Verify your instrument’s firmware is: A.13.55 or greater.
—
Verify your instrument’s IF Multiplexer (
IF MUX
) board, has P3, P203,
P403, P603, P803 connectors. If not, Keysight will purchase a new IF
MUX board. Refer to your instrument’s Service Guide, Chapter 7
bookmark “Removing and Replacing the A20 IF Multiplexer Board” in
the PDF Service Guide
a
.
—
Verify your Synthesizer board is version H or greater. If not, refer to
Appendix A:, “Synthesizer Board Upgrade (N5240-60074 (with
Tabs)/N5240-60076 (Without Tabs) Version F/G to Version H),”
a. See
“Downloading the Online PNA Service Guide” on page 7
.
1. See