N1000A Quick Start Guide

2

Installing the N1000A

signal source interferes with the calibration factors and decreases calibration 

accuracy.

4

In the dialog box, select a tab that represents the module that you want to 

calibrate.

5

Select 

Module Calibration

 and click 

Calibrate

 to starts the module calibration.

6

Follow any displayed instructions. After the calibration completes, the dialog 

box will show the status of the calibration.

Summary of Contents for N1000A DCA-X Series

Page 1: ...Keysight N1000A DCA X Series Wide Bandwidth Oscilloscope Quick Start Guide...

Page 2: ...vernment customers under its standard commercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www keysight com find sweula The license set fort...

Page 3: ...l the Modules 24 Step 5 Connect the Keyboard and Mouse optional 25 Step 6 Connect the Power 26 Step 7 Turn On the Instrument 28 Step 8 Configure a LAN Connection Optional 30 Step 9 Install a Feature L...

Page 4: ...4 N1000A Quick Start Guide Returning the Instrument for Service 41 Preparing the instrument for shipping 42...

Page 5: ...1000A 11 Accessories 12 Safety Information 14 This chapter introduces you to the N1000A and contains important safety information To install your N1000A refer to Chapter 2 Installing the N1000A The fo...

Page 6: ...ctors As with any computer you should consider the following points to maintain security and protect your N1000A Install Windows security updates as soon as they become available Consider installing a...

Page 7: ...Introduction 1 N1000A Quick Start Guide 7 Front Panel Features Figure 3 Front Panel Features...

Page 8: ...8 N1000A Quick Start Guide 1 Introduction Rear Panel Features Figure 4 Rear Panel Features...

Page 9: ...em To access the Help system complete the installation steps turn on the N1000A and click Contents on the Help menu Figure 5 N1000A with a Module Compatible Products Modules installed in module bay Al...

Page 10: ...f Connector Care Information in Help WARNING If flammable fluids are used to clean connectors the fluid shall not be placed on the instrument during use or when connected to mains voltage Cleaning the...

Page 11: ...To prevent electrical shock disconnect the Keysight Technologies Model N1000A from mains before cleaning Use a dry cloth or one slightly dampened with water to clean the external case parts Do not att...

Page 12: ...ectrical Channel Inputs SMA f f adapter Keysight part number 1250 1158 APC 3 5 f f adapter Keysight part number 5061 5311 Optional Adapters for Optical Channel Inputs Table 1 on page 13 shows the avai...

Page 13: ...was damaged contact the carrier then contact the nearest Keysight Sales Office Keep the shipping materials for the carrier s inspection The Keysight Sales Office will arrange for repair or replacement...

Page 14: ...pport WARNING If this product is not used as specified the protection provided by the equipment could be impaired This product must be used in a normal condition in which all means for protection are...

Page 15: ...instrument Another area of concern is verification that the maximum ambient operating temperature of the instrument s is not exceeded by cabinet installation Keysight recommends forced air convection...

Page 16: ...16 N1000A Quick Start Guide 1 Introduction...

Page 17: ...channel input circuits and the trigger input circuit can be damaged by electrostatic discharge ESD Therefore avoid applying static discharges to the front panel input connectors Prior to connecting a...

Page 18: ...kage Contents N1000A DCA X Series Wide Bandwidth Oscilloscope Keyboard USB Mouse USB Power cord Module bay filler panel Qty 2 ESD warning pamphlet ESD warning sticker sheet IO Libraries Media Suite CD...

Page 19: ...the 86100D were included in options Data Analytics Web Service Software Enhanced Impedance and S Parameters FlexDCA Remote Access Software FlexEye Independent Eye Acquisition and Analysis NRZ TDEC Me...

Page 20: ...ample of Static Safe Workstation WARNING These techniques for a static safe workstation should not be used when working on circuitry with a voltage potential greater than 500 volts CAUTION Electrical...

Page 21: ...aged 4 If applicable make certain that your work station air ionizer is activated and correctly positioned 5 Make sure that there are no static generators inside Keysight conductive containers with ES...

Page 22: ...itude Operating Up to 4 600 meters 15 000 ft Line Power The instruments can operate with mains supply voltage fluctuations up to 10 of the nominal voltage 100 120Vac 50 60 400 Hz 220 240Vac 50 60 Hz 7...

Page 23: ...cord is the instrument disconnecting device It disconnects the mains circuits from the mains supply before other parts of the instrument The front panel switch is only a standby switch and is not a L...

Page 24: ...ressing on the top inch of the module s front panel finger tighten the knurled locking screws Figure 8 Installing a Module DCA X DCA X Series Wide Bandwidth Oscilloscope N1000A 1 2 3 4 Trigger Input 2...

Page 25: ...nd keyboard or provide your own Since the N1000A has a touch screen the use of these devices is optional When connecting or using a mouse or keyboard always consider ergonomic principles as explained...

Page 26: ...uct damage CAUTION Do not connect ac power until you have verified the line voltage is correct as described in Table 2 on page 22 Damage to the equipment could result CAUTION This instrument has auto...

Page 27: ...rt Guide 27 NOTE The main power cord can be used as the system disconnecting device It disconnects the mains circuits from the mains supply NOTE The products can operate with mains supply voltage fluc...

Page 28: ...l be asked to accept the Windows 10 End User License Agreement EULA Once accepted the N1000A will continue to configure itself 3 After the instrument boots up the FlexDCA oscilloscope application appe...

Page 29: ...the display Instead use your finger CAUTION To ensure proper operation of the N1000A open the instrument s help system Click under Safety click Introduction and in the displayed topic click the Windo...

Page 30: ...tart the N1000A 3 Click File Exit to close the N1000A DCA application and view the Windows desktop 4 Use Windows to setup your instrument with the appropriate client protocol and configuration for you...

Page 31: ...Software Entitlement Certificate that you must use to install the license on your N1000A Refer to the instrument s help system for information on installing licenses Step 10 Connect a Printer Optiona...

Page 32: ...N1000A Step 11 Connect an External Display Optional 1 Connect your display to the rear panel Display Port or VGA connector 2 Refer to Windows Help for information on using a Windows second desktop fe...

Page 33: ...the mainframe exceeds 5 C compared to the temperature of the last vertical amplitude calibration T 5 C The time since the last calibration has exceeded 10 hours To perform a module calibration 1 Turn...

Page 34: ...ses calibration accuracy 4 In the dialog box select a tab that represents the module that you want to calibrate 5 Select Module Calibration and click Calibrate to starts the module calibration 6 Follo...

Page 35: ...35 Step 13 Learn How to Use and Program the N1000A The help system is the N1000A s user and programming manual Click Help Contents to access the help The help system s home topic should look similar t...

Page 36: ...36 N1000A Quick Start Guide 2 Installing the N1000A...

Page 37: ...does not need your hard disk to perform the calibration The hard disk includes both drive C INFINIIUM and drive D USER Drive C contains the following items Operating system N1000A application Any fact...

Page 38: ...hard disk To Declassify Instrument Memory All sensitive instrument information is written to the instrument s hard disk To declassify instrument memory Optional Press the Clear button in the N1000A s...

Page 39: ...of the line module input power The chassis ground symbol The chassis ground symbol is used to indicate a chassis connection The CE mark is a registered trademark of the European Community The CSA mark...

Page 40: ...ts a different number these topics are addressed in the addendum disclaimer statements South Korean Certification KC mark includes the marking s identifier code which follows this format MSIP REM YYY...

Page 41: ...contract it will be repaired under the terms of the warranty or contract If the instrument is no longer under warranty or is not covered by an maintenance plan Keysight will notify you of the cost of...

Page 42: ...ginal cal data disks 2 Cover all front or rear panel connectors that were originally covered when you first received the instrument 3 Pack the instrument Use original packaging or comparable Original...

Page 43: ...revent it from moving in the carton If packing foam is not available the best alternative is S D 240 Air Cap from Sealed Air Corporation Commerce California 90001 Air Cap looks like a plastic sheet fi...

Page 44: ...44 N1000A Quick Start Guide 5 Contacting Keysight...

Page 45: ...nput connector 10 instrument checking the shipment 5 9 declassify memory 38 markings 39 options 18 positioning 22 returning for service 41 ISM1 A 39 K keyboard 18 25 Keysight Technologies contacting 4...

Page 46: ...46 N1000A Quick Start Guide W warm up time 33 weight 22 Windows XP second desktop 32 setup wizard 28...

Reviews: