Keysight MIPI M-PHY Frame Generator User Guide
45
Using the Software
5
Range of Amplitude: From 200mV to 2V. Default: 400mV.
The range of High Level is defined as:
2/3 V
term
- 0.95 V < High Level < V
term
+ 2 V
The range of Low Level is defined as:
2/3
V
term
- 1 V < Low Level <
V
term
+ 1.95 V
The range of Offset is defined as:
1/2 (High Level + Low level)
Data Channel Settings
Refer to
.
•
Termination
: This is set as 100
Ω
for data channels.
•
Level Mode
: The levels can be set either in terms of High Level/Low
Level or in terms of Amplitude/Offset.
Range of Amplitude: From 35mV to 630mV. Default: 200mV.
The range of High Level is defined as:
2/3 V
term
- 0.95 V < High Level < V
term
+ 2 V
The range of Low Level is defined as:
2/3 V
term
- 1 V < Low Level < V
term
+ 1.95 V
The range of Offset is defined as:
1/2 (High Level + Low level)
•
Equalization Settings
: The equalization can be specified in terms of
De-Emphasis (Post-Cursor 1) and Preshoot (Pre-Cursor 1), in dB. The
equivalent in terms of FIR (Finite Impulse Response) Tap Coefficients is
displayed as read-only values. Note that the values of De-Emphasis
and Preshoot applied to the signal may differ slightly from what is
expected, as a result of rounding errors.
Summary of Contents for MIPI M-PHY
Page 1: ...User Guide Keysight MIPI M PHY Frame Generator ...
Page 5: ...Keysight MIPI M PHY Frame Generator User Guide 1 Introduction Overview 6 ...
Page 8: ...8 Keysight MIPI M PHY Frame Generator User Guide 1 Introduction ...
Page 18: ...18 Keysight MIPI M PHY Frame Generator User Guide 3 Software Installation and Update ...
Page 22: ...22 Keysight MIPI M PHY Frame Generator User Guide 4 Starting and Registering the Software ...
Page 70: ...70 Keysight MIPI M PHY Frame Generator User Guide 6 Software Installation and Update ...
Page 74: ...72 Keysight MIPI M PHY Frame Generator User Guide 7 Software Installation and Update ...
Page 75: ...Keysight MIPI M PHY Frame Generator User Guide 75 ...