57
Keysight InfiniiVision 2000/3000 X-Series Oscilloscope
Service Guide
3 Calibrating and Adjusting
This chapter explains how to adjust the oscilloscope for optimum operating
performance. You should perform self-calibration according to the following
recommendations:
• Every two years or after 4000 hours of operation
• If the ambient temperature is >10 °C from the calibration temperature
• If you want to maximize the measurement accuracy
The amount of use, environmental conditions, and experience with other
instruments help determine if you need shorter adjustment intervals.
Let the Equipment Warm Up Before Adjusting
Before you start the adjustments, let the oscilloscope and test equipment warm
up for at least 30 minutes.
Read All Cautions and Warnings
Read the following cautions and warning before making adjustments or
performing self-calibration.
WARNING
HAZARDOUS VOLTAGES!
Read the safety notice at the front of this book before proceeding.
Maintenance is performed with power supplied to the oscilloscope and with the protective covers
removed. Only trained service personnel who are aware of the hazards involved should perform the
maintenance. Whenever possible, perform the procedures with the power cord removed from the
oscilloscope.
CAUTION
REMOVE POWER TO AVOID DAMAGE!
Do not disconnect any cables or remove any assemblies with power applied to the oscilloscope.
Otherwise, damage to the oscilloscope can occur.
Summary of Contents for InfiniiVision 2000 X Series
Page 1: ...Keysight InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide...
Page 8: ...8 2000 3000 X Series Oscilloscopes Service Guide Contents...
Page 10: ...10 2000 3000 X Series Oscilloscopes Service Guide Figures...
Page 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Tables...
Page 72: ...72 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies...