2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
Figure 5
Setting up Equipment for Dual Cursor Accuracy Test
6
Adjust the output so that the multimeter reading displays the first Volts/div
calibrator setting value in
(depending on the oscilloscope
model).
7
Disconnect the multimeter.
8
Wait until the measurement settles.
9
Press the
Y2
softkey, then position the Y2 cursor to the center of the voltage
trace using the Entry knob.
Oscilloscope
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
BNC (f) to dual
bananna adapter