Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
27
1
Turn on the test equipment and the oscilloscope. Let them warm up for 30
minutes before starting the test.
2
Set up the oscilloscope calibrator.
a
Set the oscilloscope calibrator to provide a DC offset voltage at the
Channel 1 output.
b
Use the multimeter to monitor the oscilloscope calibrator DC output
voltage.
3
Use the 8-by-2 test connector and the BNC cable assembly to connect digital
channels D0-D7 to one side of the BNC Tee. Then connect the D0-D7 ground
lead to the ground side of the 8-by-2 connector. See
.
Table 4
Equipment Required to Test Digital Channel Threshold Accuracy
Equipment
Critical Specifications
Recommended Model/Part
Digital Multimeter
0.1 mV resolution,
0.005% accuracy
Keysight 34401A
Oscilloscope Calibrator
DC offset voltage 6.3 V
Fluke 5820A
BNC-Banana Cable
Keysight 11001-66001 or Pomona
2BC-BNC-36
BNC Tee
Keysight 1250-0781 or
Pomona 3285
50
Ω
BNC Cable
Keysight 10503A
BNC Test Connector, 8-by-2
User-built (See
Probe Cable
Keysight N6450-60001 (16-channel)
or Keysight N6459-60001 (8-channel)