2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
3
Connect the equipment (see
a
Connect the signal generator output to the oscilloscope channel 1 input.
Figure 8
Setting Up Equipment for Internal Trigger Sensitivity Test
b
With 2000 X-Series oscilloscopes, connect a 50 ohm feedthrough
termination between the channel 1 input and the BNC cable.
With 3000 X-Series oscilloscopes, set channel 1
Imped
to
50 Ohm
.
4
To verify the trigger sensitivity at the oscilloscope’s maximum bandwidth, set
the output frequency of the signal generator to the maximum bandwidth of the
oscilloscope:
•
1 GHz models: 1 GHz.
•
500 MHz models: 500 MHz.
•
350 MHz models: 350 MHz.
•
200 MHz models: 200 MHz.
•
100 MHz models: 100 MHz.
•
70 MHz models: 70 MHz.
Signal
Generator
Oscilloscope
N Cable
N to BNC Adapter