V2802A Option M81
Front and Rear Panel Features
Figure 2
V2802A-M81 Rear Panel
Chassis Ground
Line Module
Power Cords
A threaded terminal post for connecting the Distribution Unit to a conductive object,
cabinet or structure to ensure a common potential and reduce leakage current in a
system. Requires a (M4 x 0.7) threaded nut.
The line fuse, as well as a spare, reside within the line module.
illustrates
where the fuses are located and how to access them.
Install the instrument so that the ON/OFF switch is readily identifiable and is easily
reached by the operator. The ON/OFF switch is the instrument disconnecting device.
It disconnects the mains circuits from the mains supply before other parts of the
instrument, or the detachable power cord can be removed from the electrical supply.
Alternately, an externally installed switch or circuit breaker which is readily
identifiable and is easily reached by the operator may be used as a disconnecting
device.
A line power cord is supplied in one of several configurations, depending on the
destination of the original shipment. Keysight can supply additional certified power
cords to meet region electrical supply and receptacle configurations. Please contact
Keysight at:
www.keysight.com
for assistance in power cord selection.
This is a Safety Class I Product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall only be inserted in a socket
outlet provided with a protective earth contact. Any interruption of the protective
conductor inside or outside of the product is likely to make the product dangerous.
Intentional interruption is prohibited.
Chassis Ground
Line Module