Safety and Regulatory Information
Before Applying Power
Verify that the premises electrical supply is within the range of the instrument. The
instrument has an autoranging power supply.
If this product is not used as specified, the protection provided by the equipment
could be impaired. This product must be used in a normal condition (in which all
means for protection are intact) only.
The Mains wiring and connectors shall be compatible with the connector used in the
premise electrical system. Failure, to ensure adequate earth grounding by not using
the correct components may cause product damage, and serious injury.
This product is designed for use in Installation Category II and Pollution Degree 2.
Verify that the premise electrical voltage supply is within the range specified on the
instrument.
When installing the product in a cabinet, the convection into and out of the product
must not be restricted. The ambient temperature (outside the cabinet) must be less
than the maximum operating temperature of the product by 4 °C for every 100 watts
dissipated in the cabinet. If the total power dissipated in the cabinet is greater than
800 watts, then forced convection must be used.
This instrument has auto-ranging line voltage input, be sure the supply voltage is
within the specified range and voltage fluctuations do not to exceed 10 percent of
the nominal supply voltage.