U3042AM04
Controlling the Test Set with
N5222/32/42A/B
2.
On the analyzer select Response > Cal Wizard
-OR-
(Response > Cal > Other Cals...)
.
a.
If using a mechanical cal kit, select SmartCal (Guided Calibration) > Next.
b.
If using an ECal module, connect the ECal to an available USB port and select
Use Electronic Calibration (ECal) > Next
-OR-
(Response > Cal > Other Cals > ECal...)
.
3.
On the “Select Calibration Ports and ECal Module” dialog box press Select All, or
select the ports you are calibrating and press Next. If an ECal characterization has
been done, select
V
iew/Select ECal Module and select the previously saved user
file
from
step 1
.
Figure 3
2
8
-Port Calibration
4. Connect the ECal or the mechanical cal kit to the ports you are calibrating
following the Cal Wizard prompts and click Measure after each connection.
5. At the Calibration Completed prompt, select Save As User Calset and enter the
name desired and Save.
6. After calibrating test set ports, use a quality load and short to verify the
calibration on each port or end of the test cable. Measure reflection and confirm
the return loss is as expected. If the result is not as expected, repeat the
calibration without the test set and ensure that the analyzer is in standard
(non-multiport) mode.
If measurement errors occur, ensure the newest version of firmware is installed on
the analyzer. Measurement errors can be a result of firmware algorithms. Consult
with Keysight Service or firmware web page for the latest firmware revisions and
history at
http://na.support.keysight.com/pna/firmware/firmware.htm
.
Help
Keysight U3042AM04 User's and Service Guide
32