U3042AM04
System Setup with
N5222/42A/B
N5222
/42A/B
RF Interface Cable Connections
illustrates the cable configuration of the Test Set to the
analyzer. The cables have been supplied in the interface cable kit (U3021-60047).
1.
Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM jumpers
(x16) on the PNA. The RCVR R1 to SOURCE OUT reference loop jumper (x1)
remain
on the front panel.
2.
Connect the RF interconnect cables from the analyzer to the Test Set in the order
listed. As you are connecting each cable, torque to 8 in-lb. The longer, straight end
of each cable is connected to the Test Set. Refer to
and
Over torque will cause damage to the Test Set and may cause connectors to spin or
become loose.
Table
9
PNA or PNA-X RF Interface Cable Connection (U3021-60047)
Numeric
Order
RF Cables
From PNA-X
To Test Set
1
U3042-20031
Port 1 SOURCE OUT Port 5 SOURCE IN
2
U3042-20032
Port 1 CPLR THRU
Port 5 CPLR THRU
3
U3042-20029
Port 1 CPLR ARM
Port 5 CPLR ARM
4
U3042-20030
Port 1 RCVR A IN
Port 5 RCVR OUT
5
U3042-20035
Port 3 SOURCE OUT
Port 6 SOURCE IN
6
U3042-20033
Port 3 CPLR ARM
Port 6 CPLR ARM
7
U3042-20034
Port 3 RCVR C IN
Port 6 RCVR OUT
8
U3042-20036
Port 3 CPLR THRU
Port 6 CPLR THRU
9
U3042-20039
Port 4 SOURCE OUT Port 7 SOURCE IN
10
U3042-20037
Port 4 CPLR ARM
Port 7 CPLR ARM
11
U3042-20038
Port 4 RCVR D IN
Port 7 RCVR OUT
12
U3042-20040
Port 4 CPLR THRU
Port 7 CPLR THRU
13
U3042-20041
Port 2 CPLR ARM
Port 8 CPLR ARM
14
U3042-20044
Port 2 CPLR THRU
Port 8 CPLR THRU
15
U3042-20043
Port 2 SOURCE OUT
Port 8 SOURCE IN
16
U3042-20042
Port 2 RCVR B IN
Port 8 RCVR OUT
Keysight U3042AM04 User's and Service Guide
20