U3042AM04
Controlling the Test Set with
N5222/32/42A/B
RF Path Configuration with Analyzer Option 029
If the analyzer has Option 029 or H29 (noise figure capability) verify that the path
configuration is set to Default. If it is not in Default configuration, Port 5 will not
operate correctly.
1.
Select Trace/Chan > Channel > Hardware Setup > Path Cong...
-OR-
(Instrument > Setup > Internal Hardware > RF Path Config...)
a
nd in the drop-
down menu select Default > OK. Save this configuration as a User Preset by
selecting Save > User Preset > Save current state as User Preset. Do not use the
factory Preset (User Preset Off), the analyzer will return to Option 029 path
configuration.
2.
Verify that the Port 1 Noise Tuner Switch is set to external. Select
Utilities System>
C
onfigure > Preferences
-OR-
(Utility > System > System Setup > Preferences > User Preset)
.
If not, select Meas: Port 1 Noise Tuner Switch is set to
external.
Figure 3
0
RF Path Configuration
Keysight U3042AM04 User's and Service Guide