Key Reference
Mode
79
Toggles the zeroing type of the Power Sensor
between internal and external. When this type is set
to EXT, pressing Zeroing will bring up a message
which indicates the external signal need be pulled
out. This setting is not affected by the preset in Power
Meter menu.
Key access:
[MODE] > {Power Meter} > {Zero}
Meas Disp
Pressing this key accesses the Meas Disp submenu for
defining the display range and resolution while in Power
Meter mode.
• Disp Range
Access the associated Disp Range submenu.
•
Auto Range
Toggles the auto range function between On and Off.
When the Auto Range is set to On, the Top and Bot-
tom value of the power meter’s display range will be
set automatically according to the measurement
data:
Top = Max. data + 0.5
Bottom = Min. data - 0.5
The default setting is Off.
•
Top
Set the top value of the power meter’ display range.
Range from ( 0.01) to 200 dBm.
The default value is 20 dBm.
•
Bottom
Set the bottom value of the power meter’ display
range.
Range from -200 to (Top - 0.01) dBm.
The default value is -70 dBm.
• Resolution
Access the associated Resolution submenu.