Performance Verification
6
N2750/1/2A User’s Guide
67
NOTE
Allow the probe to warm up for at least 20 minutes.
f
If available, use the N2787A 3D Probe Positioner to position the probe
straight up and down (perpendicular) to the PV Fixture as shown in
.
Figure 46
Measurement setup for V
in
17
Spread the probe tip’s wires so that the tips are slightly wider than the gap