Probe Calibration
3
N2750/1/2A User’s Guide
41
• Clamp the probe leads to the PV fixture’s signal or ground. Position the probe
tips on the PV fixture as shown in
. When using the solder-in tip, use
the fixture’s spring-loaded clamps to hold the leads.
Figure 21
Example of Proper Lead Placement on an E2655B PV Fixture (Browser Tip
Shown)
• On 90000 Q/X-Series oscilloscopes, use an N5442A adapter when instructed to
connect the probe to one of the oscilloscope’s channel inputs.
CAUTION
To avoid damaging the oscilloscope’s
Cal Out
connector, do not apply
force to the PV fixture. Light probe contact is all that is needed for the
calibration.
CAUTION
NEVER solder a probe tip to the thick-film gold. The gold will
immediately dissolve into the solder and disappear.
NOTE
If you are using the browser tip, it is recommended that you use the N2787A
3D Probe Positioner to hold the probe in place.
NOTE
You can check that the probe leads are correctly connected by pressing the
front panel autoscale button. A stable step should be shown on the screen.
Pressing autoscale will close the Probe Calibration dialog box.
Summary of Contents for N2750A
Page 1: ...Keysight N2750 1 2A Differential Probes User s Guide ...
Page 28: ...28 N2750 1 2A User s Guide 1 Using the Probes ...
Page 34: ...34 N2750 1 2A User s Guide 2 SPICE Models Figure 16 SPICE Model ...
Page 38: ...38 N2750 1 2A User s Guide 2 SPICE Models ...
Page 42: ...42 N2750 1 2A User s Guide 3 Probe Calibration ...
Page 54: ...54 N2750 1 2A User s Guide 5 Performance Data Plots Figure 35 Frequency Response Common Mode ...
Page 56: ...56 N2750 1 2A User s Guide 5 Performance Data Plots Figure 38 Frequency Response Common Mode ...
Page 74: ...74 N2750 1 2A User s Guide 6 Performance Verification ...